The flexibility of the Rotem RAM R-200 can be expanded with a choice of nine external interchangeable detectors
suitable for alpha, beta, low or high-energy gamma and X-ray detection.
All detectors feature:
§ Simple plug-in dual RS-232 interface
§ Automatic detector identification
§ Individual detector calibration factor stored within probe
§ Automatic unit selection according to detector
§ Continuous monitoring of accumulated dose by internal detector
§ Automatic detector self-diagnostic routines
§ Detectors can also be used in a stand-alone mode with Rotem software packages
Page 1
|
Radiation Detected: |
Gamma above 50 keV |
|
Scintillator: |
NaI(Tl) 2" diameter, 2" thick. Window of 1 mm. (0.04") Al. |
|
Sensitivity: 137Cs: |
20,000
cps/mR/h |
|
Energy Calibration: |
Software calibrated single channel analyzer (SCA). |
|
|
0 to 50,000cps |
|
Output Signals: |
RS-232
or RS- 485 and TTL pulses. |
|
|
-30oC to +60oC (-22oF to +140oF) |
|
|
10% to 95% RH (non condensing) |
|
Dimensions: |
340mm
(13.4") length, 70mm (2.75") Diam. |
|
Weight: |
1.75kg (3.9lbs) |
|
Casing: |
Aluminum, splash proof, IP-65 |
|
Max. Cable Length: |
TTL pulse - 100 m, RS-232 - 12 m, RS-485 - 1.2 km |
|
0 – 1.00 Mcpm |
Range: |
|
1.5 - 2.0 mg/cm2 |
Window: |
|
6.4 cm2 |
Effective area: |
|
1500 cpm/mR/h (60CO) |
Sensitivity: |
|
LND 7231 |
Geiger: |
|
±
10% of reading |
Accuracy: |
|
-300C to +600C |
|
|
280 gr. (0.6 lb) |
Weight: |
|
40 mm (1.6”) |
Diameter: |
|
155 mm (6.1”) |
Length: |
|
Type: |
Air proportional |
|
Window Thickness: |
0.85mg/cm2 aluminized window |
|
Effective area: |
100 cm2 (15.5 in2) |
|
Probe Efficiency (2p): |
*18% |
|
Surface
Sensitivity: |
*530 cpm/Bq/cm2compensated in the meter to 100% to be 3000 cpm/Bq/cm2 |
|
Range: |
0 to 1.0 Mcpm |
|
Accuracy: |
* ±15% of reading |
|
|
-100C
to +500C |
|
Weight: |
1Kg (2.2 lbs) |
|
Width: |
75mm (3") |
|
Height: |
55mm (2.2") |
|
Length: |
240mm (9.5") |
Page 1