The flexibility of the Rotem RAM DA Series can be expanded with a choice of nine external interchangeable detectors
suitable for alpha, beta, low or high-energy gamma and X-ray detection.
All detectors feature:
§ Simple plug-in digital interface
§ Automatic detector identification
§ Individual detector calibration factor stored within probe
§ Automatic unit selection according to detector
§ Optional extended cable lengths
§ Automatic detector self-diagnostic routines
Application: Measurement of gamma radiation over a very wide range. Based on different types of energy compensated GM tubes.Extended hook-up cable of up to 100 meters can be used.
Measuring Range:
GM-40: 250 µSv/h to 10 Sv/h
GM-41: 50 µSv/h to 1 Sv/h
GM-42:
0.5 µSv/h to 10 mSv/h
Application: Efficient surface monitoring
of alpha, beta and gamma
radiation contamination,
based on 1ľ” Pancake GM
tube.
Sensitivity(137Cs):
~350 cpm/µSv/h
(3500 cpm/mR/h)
Measuring Range:
0 to 42,000 cps
Application: Surface monitoring for Alpha contamination.
Based on Ionization chamber.
Very easy to maintain.
100 cm2 (15.5 sq.
in.)
Surface Sensitivity:
530 cpm per Bq/cm2
Measuring Range:
0 to 50,000 cps
Application: Wide range gamma, x-ray and beta radiation measurement.
Based on 0.5 liter atmospheric
ionization chamber.
Uses special
features and
advanced technology.
Measuring Range:
1.0 µSv/h to 250 mSv/h
Application: Gamma scintillation probe for improved x-ray and low
energy gamma radiation
detection,based on thin Nal 2”
diameter
scintillator
(2” dia. X 0.04” thickness).
S.C.A. can be set within the
energy range, optimized for 125I
detection and/or other isotopes
Energy range: 10 to 80 keV
125I Surface sens. : 440 cpm/Bqcm2
Measuring range: 0 to 50,000 cps
detection. Nal 2” x 2” diameter scintillator.Optional Single
channel analyzer (SCA)
Surface Sensitivity (in contact):
131I: 320 cpm/Bq/cm2
99mTc: 315 cpm/Bq/cm2
Range: 0 to 50,000 cps